A spectrophotometer is an instrument
used to measure the intensity

00:00:12.387 --> 00:00:15.598
of light at a particular wavelength
in the visible spectrum.

00:00:16.266 --> 00:00:19.310
This type of instrument can be used
in the clinical laboratory

00:00:19.310 --> 00:00:23.106
to determine the concentration
of a substance in solution, by

00:00:23.106 --> 00:00:26.735
measuring the amount of light
either transmitted or absorbed.

00:00:27.068 --> 00:00:31.364
When a beam of light is passed through a
sample reaction at a specific wavelength.

00:00:31.990 --> 00:00:34.659
For this video,
we will be using a single beam

00:00:34.659 --> 00:00:37.662
spectrophotometer with a digital display.

00:00:38.246 --> 00:00:41.458
This instrument has a wavelength
range of 340

00:00:41.458 --> 00:00:46.629
nanometers to 950 nanometers,
and a nominal spectral bandwidth

00:00:46.880 --> 00:00:49.591
of 20 nanometers.

00:00:49.591 --> 00:00:53.219
The main controls for routine operation
of this spectrophotometer

00:00:53.386 --> 00:00:57.307
include the power switch, zero control

00:00:58.058 --> 00:01:02.353
percentage T,
the transmittance absorbance control,

00:01:03.188 --> 00:01:06.733
100% T 0.0 A,

00:01:07.609 --> 00:01:11.654
the wavelength control,
and the mode selection button.

00:01:13.073 --> 00:01:13.615
The power

00:01:13.615 --> 00:01:18.495
Switch zero control knob is used to power
the instrument on and off,

00:01:18.703 --> 00:01:22.165
and to set the digital display
to 0% transmittance

00:01:22.499 --> 00:01:27.295
or T when the sample compartment is empty
and the cover is closed.

00:01:28.421 --> 00:01:33.218
The transmittance absorbance control
knob is used to set the digital display

00:01:33.218 --> 00:01:38.890
to 100% transmittance T or 0.0 absorbance

00:01:39.265 --> 00:01:43.561
A when a water blank or reagent
blank is inserted

00:01:43.561 --> 00:01:46.564
in the sample compartment
and the cover is closed.

00:01:47.565 --> 00:01:50.568
The wavelength control
knob is used to select

00:01:50.568 --> 00:01:53.571
the desired wavelength for the analysis.

00:01:53.947 --> 00:01:58.159
The mode selection button
is used to select between transmittance,

00:01:58.409 --> 00:02:02.413
absorbance, concentration, or factor mode.

00:02:03.289 --> 00:02:07.544
For most analysis, the absorbance
or transmittance mode will be used.

00:02:08.044 --> 00:02:11.256
Some spectrophotometer
may also have a filter level

00:02:11.256 --> 00:02:14.259
that must be adjusted to select
the appropriate filter,

00:02:14.300 --> 00:02:19.347
based on the wavelength to be used
to power on the spectrophotometer.

00:02:19.681 --> 00:02:22.600
Plug the power cord
into an appropriate outlet

00:02:22.600 --> 00:02:25.645
and rotate the power switch zero control

00:02:26.020 --> 00:02:29.357
percentage knob in a clockwise direction.

00:02:30.692 --> 00:02:33.236
When working
with instruments or equipment,

00:02:33.236 --> 00:02:37.365
always check the power cords for damage
before inserting into a power source.

00:02:38.366 --> 00:02:41.244
The instrument will click on
and the display will light up.

00:02:41.244 --> 00:02:44.205
Once the instrument has power.

00:02:44.205 --> 00:02:46.124
The instrument
should be allowed to warm up

00:02:46.124 --> 00:02:49.085
for at least 15 minutes prior to use.

00:02:52.213 --> 00:02:54.257
After the instrument is warmed up,

00:02:54.257 --> 00:02:58.136
set the desired wavelength
by turning the wavelength control knob.

00:02:59.262 --> 00:03:03.641
Set the filter lever to the appropriate
position for the selected wavelength.

00:03:03.975 --> 00:03:05.143
For example,

00:03:05.143 --> 00:03:08.980
if the desired wavelength for testing
is 450 nanometers,

00:03:09.480 --> 00:03:11.608
the lever should be positioned
to the left,

00:03:11.608 --> 00:03:16.404
so the 340 to 599 nanometer
filter is in place.

00:03:17.113 --> 00:03:21.284
If the required wavelength were instead
650 nanometers,

00:03:21.701 --> 00:03:23.828
the lever
should be positioned to the right,

00:03:23.828 --> 00:03:28.208
so the 600 to 950
nanometer filter is in place

00:03:29.626 --> 00:03:30.668
to account for any

00:03:30.668 --> 00:03:34.130
stray light in the sample compartment
that could affect measurements.

00:03:34.631 --> 00:03:37.508
The instrument is first blanked or zeroed

00:03:37.508 --> 00:03:40.511
with nothing in the sample compartment

00:03:41.596 --> 00:03:43.765
to zero the spectrophotometer.

00:03:43.765 --> 00:03:47.518
Ensure that the sample compartment is
empty and the cover is closed.

00:03:48.269 --> 00:03:51.981
Turn the power switch zero control
knob percentage T

00:03:52.607 --> 00:03:56.653
until the digital display shows
the transmittance as 0%.

00:03:57.487 --> 00:04:00.615
Please note that this may require several
turns of the knob.

00:04:01.449 --> 00:04:04.118
Some procedures require
that the transmittance

00:04:04.118 --> 00:04:07.121
is monitored
while others measure absorbance.

00:04:07.538 --> 00:04:10.917
Refer to your procedure
to identify the desired measurement,

00:04:11.251 --> 00:04:14.796
and use the mode selection button
to toggle to the appropriate setting.

00:04:15.463 --> 00:04:19.801
Most procedures also require
the instrument to be zeroed using a water

00:04:19.801 --> 00:04:23.805
blank or reagent blank, to account
for the effects of the solution itself.

00:04:24.430 --> 00:04:27.767
To prepare a water blank,
fill a clean cuvette

00:04:27.767 --> 00:04:30.853
with distilled water,
avoiding the formation of bubbles.

00:04:31.688 --> 00:04:33.940
If using a reagent blank, fill

00:04:33.940 --> 00:04:36.943
a clean cuvette
with the designated reagent.

00:04:37.443 --> 00:04:40.947
In general, corvettes should be filled
to at least half their volume.

00:04:41.489 --> 00:04:45.994
The minimum required volume for the
spectrophotometer is three milliliters.

00:04:46.661 --> 00:04:50.456
Using a lint free tissue, gently wipe
the outside of the cuvette

00:04:50.498 --> 00:04:55.753
to remove any surface interference,
such as fingerprints, dust, or liquid.

00:04:56.462 --> 00:04:59.841
Insert the cuvette
into the sample compartment and align

00:04:59.841 --> 00:05:03.303
the guide mark on the cuvette
with the guide mark on the compartment.

00:05:04.012 --> 00:05:09.100
Gently but firmly push the cuvette into the sample compartment and close the cover.

00:05:10.601 --> 00:05:13.604
Using the transmittance absorbance
control knob

00:05:13.855 --> 00:05:17.317
100% T 0.0 A.

00:05:18.067 --> 00:05:21.988
Adjust the display to 100% transmittance T

00:05:22.530 --> 00:05:25.450
or zero absorbance A

00:05:25.450 --> 00:05:28.328
depending on the mode previously selected.

00:05:28.328 --> 00:05:31.956
Remove the cuvette and appropriately
discard the water reagent.

00:05:32.665 --> 00:05:35.668
Do not discard the cuvette as it will be
reused.

00:05:36.085 --> 00:05:39.797
Rinse the cuvette three times
with distilled water and gently tap

00:05:39.797 --> 00:05:43.176
the rinsed cuvette
on a small stack of paper towels to remove

00:05:43.176 --> 00:05:44.886
the remaining water.

00:05:44.886 --> 00:05:48.014
Use a lint free tissue to wipe
the outside of the cuvette.

00:05:48.890 --> 00:05:52.518
To measure the transmittance
or absorbance of a sample reaction.

00:05:52.935 --> 00:05:55.688
Fill a clean cuvette
with at least three milliliters

00:05:55.688 --> 00:05:58.691
of the solution,
avoiding formation of any bubbles.

00:05:59.359 --> 00:06:03.404
Use a lint free tissue to wipe
the outside of the cuvette and insert

00:06:03.404 --> 00:06:07.533
the cuvette into the sample compartment,
making sure to align the guide marks.

00:06:08.534 --> 00:06:09.786
Close the cover

00:06:09.786 --> 00:06:14.624
and record the percentage transmittance
or absorbance from the digital display.

00:06:15.041 --> 00:06:18.461
Remove the cuvette and appropriately
discard the sample reaction.

00:06:19.837 --> 00:06:21.631
Rinse the cuvette three times

00:06:21.631 --> 00:06:24.675
with distilled water, and gently tap
the rinsed cuvette

00:06:24.675 --> 00:06:28.096
on a small stack of paper towels
to remove the remaining water.

00:06:29.347 --> 00:06:32.475
Use a lint free tissue to wipe
the outside of the cuvette.

00:06:33.434 --> 00:06:36.938
Repeat this process
for all remaining samples to be measured.

00:06:38.022 --> 00:06:41.609
Once all measurements are complete,
the spectrophotometer

00:06:41.609 --> 00:06:44.654
should be powered off
by turning the power switch

00:06:44.654 --> 00:06:49.117
zero control
knob percentage counterclockwise

00:06:49.117 --> 00:06:52.120
until the instrument clicks
and the display is blank.

00:06:52.912 --> 00:06:55.915
Carefully
unplug the power cord from the outlet.

00:06:56.124 --> 00:06:59.377
Clean up any spills on the instrument
using a lint free tissue

00:06:59.585 --> 00:07:03.756
dampened with water or ethanol,
followed by a paper towel dampened

00:07:03.756 --> 00:07:07.927
with plain water or around the instrument
with a bench cleaning solution.

00:07:08.803 --> 00:07:12.348
Completely dry the exterior
of the instrument with a lint free wipe.

00:07:13.182 --> 00:07:17.895
The spectrophotometer should be covered
and placed in a secure area for storage.