﻿WEBVTT

00:00:08.108 --> 00:00:09.309
Mass spectrometry

00:00:09.309 --> 00:00:13.246
allows for the identification
and quantitation of specific analytes

00:00:13.246 --> 00:00:17.550
with greater sensitivity and specificity
than many other lab techniques.

00:00:18.752 --> 00:00:21.721
Samples are brought to the lab,
sorted by test.

00:00:29.596 --> 00:00:33.967
The analyte of interest is then extracted
from the samples submitted to the lab,

00:00:33.967 --> 00:00:37.370
along with collaborators
and controls to be used on the run.

00:00:42.242 --> 00:00:44.544
Quantitation
of an unknown amount of analyte

00:00:44.544 --> 00:00:49.149
in each sample by mass spec is achieved
using calibrator and an internal standard.

00:00:49.549 --> 00:00:53.019
The acceptability of each run is evaluated
with different levels

00:00:53.019 --> 00:00:56.022
of controls.

00:00:57.891 --> 00:01:00.960
Collaborators and controls are samples
that have a known amount

00:01:00.960 --> 00:01:04.764
of the analyte added to a matrix
that matches the matrix of the patient.

00:01:04.764 --> 00:01:05.832
Samples.

00:01:05.832 --> 00:01:08.635
A fixed amount of an internal
standard is added

00:01:08.635 --> 00:01:11.638
to the collaborators
controls and patient samples

00:01:11.704 --> 00:01:15.208
to allow the amount of analyte
in a patient sample to be quantified.

00:01:16.009 --> 00:01:19.012
The internal standard is typically either
an isotope

00:01:19.045 --> 00:01:22.082
labeled version of the analyte,
or another compound

00:01:22.082 --> 00:01:25.351
with very similar chemical characteristics
as the analyte.

00:01:26.119 --> 00:01:29.456
The internal standard helps
to account for day to day variations

00:01:29.456 --> 00:01:32.959
observed during the extraction
and analysis of the samples.

00:01:34.294 --> 00:01:34.928
The samples

00:01:34.928 --> 00:01:38.198
are further processed
using one of several techniques available.

00:01:38.832 --> 00:01:42.535
Labs determine the most suitable process
to use based on the chemistry

00:01:42.535 --> 00:01:45.538
and matrix of the analytes being tested.

00:01:46.639 --> 00:01:47.807
The sample of interest

00:01:47.807 --> 00:01:51.778
is extracted using column chemistry,
where impurities and other interfering

00:01:51.778 --> 00:01:54.781
substances are removed
from the original specimen.

00:01:58.785 --> 00:02:00.086
The specimen is transferred

00:02:00.086 --> 00:02:03.089
to a 96 well plate for the next step.

00:02:03.723 --> 00:02:06.192
The samples are
then loaded into an auto sampler

00:02:06.192 --> 00:02:09.195
for instrumental analysis.

00:02:29.616 --> 00:02:30.750
For clinical testing.

00:02:30.750 --> 00:02:33.753
There are three types of mass spec
analysis in common use.

00:02:34.687 --> 00:02:35.889
Liquid chromatography.

00:02:35.889 --> 00:02:39.559
Mass spec LCMs gas chromatography.

00:02:39.559 --> 00:02:42.562
Mass spec GCMs,

00:02:42.629 --> 00:02:46.266
and inductively coupled plasma
mass spec ICP.

00:02:46.266 --> 00:02:48.201
Mass.

00:02:48.201 --> 00:02:51.204
This video will show the LCMs workflow.

00:02:54.541 --> 00:02:55.408
The auto sampler

00:02:55.408 --> 00:02:58.711
introduces each sample
based on a list into the instrument,

00:02:58.711 --> 00:03:02.315
with an automated sample needle,
which is washed after each sample.

00:03:03.183 --> 00:03:07.320
The LCMs then further separates
the analyte from other sample constituents

00:03:07.320 --> 00:03:08.454
using liquid mobile

00:03:08.454 --> 00:03:12.158
phase and a stationary phase housed
within the analytical column.

00:03:12.392 --> 00:03:16.462
The analyte is introduced into a stream
of mobile phases in a specific amount

00:03:16.462 --> 00:03:20.667
that ensures the analyte adheres to the
stationary phase in the analytical column.

00:03:21.234 --> 00:03:24.971
The ratio of mobile phases is
then changed to release the analyte

00:03:24.971 --> 00:03:25.939
in a controlled manner,

00:03:25.939 --> 00:03:29.475
to separate it from any interferences
still present in the sample.

00:03:30.276 --> 00:03:33.746
The analyte is released from the column
and is carried by the mobile phases

00:03:33.746 --> 00:03:35.114
into the mass spec source.

00:03:36.082 --> 00:03:36.449
Here, the

00:03:36.449 --> 00:03:39.552
liquid stream flows into a nebulizer
where a carefully controlled

00:03:39.552 --> 00:03:43.623
combination of vacuum temperature,
gas flow, and electric voltages

00:03:43.823 --> 00:03:47.460
help vaporize the stream and introduce
a charge to the analyte of interest.

00:03:48.261 --> 00:03:51.564
This causes the solvent of the sample
to evaporate and the analyte

00:03:51.564 --> 00:03:55.501
to create an ion, which will then
enter the ion path of the instrument.

00:03:56.369 --> 00:04:00.673
Within a triple clad mass spec,
the molecular ions pass through a region

00:04:00.673 --> 00:04:04.544
where the analyte is separated further
from any mobile phase remnants.

00:04:05.411 --> 00:04:07.947
Next, the ion optics region helps

00:04:07.947 --> 00:04:11.651
focus the ion beam and three quadruples
Q and three.

00:04:11.918 --> 00:04:16.389
Further, select for the iron or analyte
of interest using electromagnetic fields,

00:04:16.623 --> 00:04:19.726
electric currents,
and an inert collision gas.

00:04:20.293 --> 00:04:24.430
Quadruples are the functional units
of a triple quad mass spectrometer.

00:04:25.098 --> 00:04:27.367
Each of these
is basically an electromagnet.

00:04:27.367 --> 00:04:28.801
In two planes.

00:04:28.801 --> 00:04:33.039
It creates a magnetic field that attracts
the ionized molecule to one of the poles,

00:04:33.373 --> 00:04:36.709
and then switches the plane and pole
to which the ion is attracted.

00:04:37.143 --> 00:04:39.712
The switching from pole to pole
is carefully controlled

00:04:39.712 --> 00:04:43.983
to cause the molecule to fly in a spiral
path down the middle of the quadrupole.

00:04:44.550 --> 00:04:47.687
The instrument is calibrated
to ensure the accurate selection

00:04:47.687 --> 00:04:52.325
of only the ions of interest
through the center of the quadrupole ions,

00:04:52.325 --> 00:04:55.561
with different mass to charge
ratios are filtered out

00:04:56.195 --> 00:04:59.432
due to the high rate of change
and the currents of the magnetic fields.

00:04:59.732 --> 00:05:02.935
Many different
analytes can be evaluated in the same run.

00:05:03.903 --> 00:05:04.504
Ions that

00:05:04.504 --> 00:05:08.541
make it through all three quadruples
reach the detector, which then amplifies

00:05:08.541 --> 00:05:11.644
the signal from the ions
to generate a voltage that is measured

00:05:11.644 --> 00:05:14.647
and recorded by the onboard computer.

00:05:14.714 --> 00:05:18.151
These voltages are then plotted on a graph
known as the Chromatogram.

00:05:18.151 --> 00:05:19.786
For each channel.

00:05:19.786 --> 00:05:23.323
Each channel represents
one precursor to product mass transition.

00:05:23.790 --> 00:05:26.626
These different transitions for each
analyte are referred

00:05:26.626 --> 00:05:30.330
to as quantifier
and qualifier ions or transitions.

00:05:30.897 --> 00:05:35.268
It is best practice to monitor two
diagnostic transitions for each analyte.

00:05:36.369 --> 00:05:39.272
The filtering, fragmenting, and filtering
again, is

00:05:39.272 --> 00:05:42.809
what makes the mass spec methods highly
specific and sensitive.

00:05:44.610 --> 00:05:47.313
Data analysts review
every run's curve controls

00:05:47.313 --> 00:05:50.850
and internal standards to ensure
all are within established parameters.

00:05:51.351 --> 00:05:55.021
All data analyzes are performed
in duplicate before releasing results

00:05:55.021 --> 00:05:55.822
to physicians.
