﻿WEBVTT

00:00:07.257 --> 00:00:09.926
After a blood specimen has been diluted
as directed

00:00:09.926 --> 00:00:13.555
by manufacturer instructions,
and the vial has been mixed.

00:00:13.805 --> 00:00:17.684
A small amount of specimen is loaded
into the hemostats chamber for counting.

00:00:18.435 --> 00:00:23.231
Materials needed for this process
include Hema cytometer with cover glass,

00:00:23.606 --> 00:00:27.152
diluted specimen
alcohol, and lint free tissues,

00:00:27.736 --> 00:00:31.573
timer, moist chamber, and a microscope

00:00:31.573 --> 00:00:34.576
with ten x and 40 x objectives.

00:00:35.827 --> 00:00:39.039
The chamber is composed of the hemostats
motor and cover glass.

00:00:39.581 --> 00:00:43.376
This cover glass is specific
for the chamber and is optically flat

00:00:43.376 --> 00:00:46.963
to ensure a consistent depth
over the entire counting chamber.

00:00:47.297 --> 00:00:50.383
The center section of the hemostats
emitter has three components

00:00:50.675 --> 00:00:53.678
the platforms, the moat, and the bridges.

00:00:53.970 --> 00:00:57.223
There are two platforms,
one on each side of the chamber.

00:00:57.766 --> 00:01:01.102
As you look closely at the platforms,
you should be able to see a grid on

00:01:01.102 --> 00:01:01.811
each side.

00:01:02.896 --> 00:01:04.522
There is a V shaped groove on

00:01:04.522 --> 00:01:07.859
each side that is used to help load
the specimen into the chamber.

00:01:08.526 --> 00:01:12.405
Around each platform is a moat which
separates each platform from the other.

00:01:13.865 --> 00:01:16.868
Beside
the moat on each side are the bridges.

00:01:17.035 --> 00:01:20.830
These raised sides provide support
for the cover glass and the distance

00:01:20.830 --> 00:01:24.375
between the platforms,
and the covered glass is 0.1mm.

00:01:24.876 --> 00:01:28.755
This space will be filled with the fluid
on which the cell count will be performed.

00:01:30.131 --> 00:01:31.966
Accurate cell counts depend

00:01:31.966 --> 00:01:35.678
on the quality of the equipment
used as well as using proper technique.

00:01:36.221 --> 00:01:38.681
Each platform is etched with a large grid

00:01:38.681 --> 00:01:41.643
measuring three millimeters on each side.

00:01:41.643 --> 00:01:45.563
This large group is subdivided into nine
one millimeter squares,

00:01:46.231 --> 00:01:50.527
each square subdivided into various
additional squares for additional utility.

00:01:51.736 --> 00:01:52.695
The first step

00:01:52.695 --> 00:01:56.825
to correctly loading the hem estimators
to ensure that it is immaculately clean.

00:01:57.575 --> 00:02:01.162
Any debris left can negatively impact
the optical quality

00:02:01.162 --> 00:02:04.332
of the chamber and cause incorrect results
to be obtained.

00:02:04.999 --> 00:02:09.045
The chamber is cleaned using isopropyl
alcohol and lint free tissue.

00:02:09.921 --> 00:02:12.924
The cover glass should always be handled
by its edges.

00:02:14.801 --> 00:02:16.219
To begin loading the chamber.

00:02:16.219 --> 00:02:19.389
The cover glass is placed on the Hema
cytometer, so both sides of

00:02:19.389 --> 00:02:22.392
the chamber are covered equally.

00:02:22.600 --> 00:02:26.020
Next, the diluted specimen is mixed gently
but thoroughly,

00:02:26.312 --> 00:02:29.399
and the first few drops of liquid
are expressed into a tissue

00:02:30.150 --> 00:02:32.861
while maintaining
pressure on the specimen vial.

00:02:32.861 --> 00:02:36.322
The next drop is applied to the V-shaped
groove at the junction of the covered

00:02:36.322 --> 00:02:37.824
glass and chamber.

00:02:37.824 --> 00:02:41.077
Pressure is continuously applied
to the vial, and the pipette is held

00:02:41.369 --> 00:02:43.454
at a 45 degree angle.

00:02:43.454 --> 00:02:46.958
The slight pressure on the vial will work
with the capillary action of the glass

00:02:46.958 --> 00:02:48.418
to properly fill the chamber.

00:02:50.295 --> 00:02:52.672
Watch carefully
and remove the pipette from the chamber

00:02:52.672 --> 00:02:55.675
when it is completely filled,
but not overfilled.

00:02:56.467 --> 00:02:59.304
If fluid flows into the mold,
the chamber is overfilled

00:02:59.304 --> 00:03:01.306
and the chamber will have to be cleaned
in the specimen.

00:03:01.306 --> 00:03:03.057
Reloaded.

00:03:03.057 --> 00:03:06.060
If it will be several minutes
before counting can occur,

00:03:06.102 --> 00:03:09.105
the chamber should be placed
in a moist chamber.

00:03:09.105 --> 00:03:13.484
This can be created by placing a damp
filter paper in an empty Petri plate,

00:03:13.693 --> 00:03:16.696
raised on pieces of wooden
applicator sticks.

00:03:17.363 --> 00:03:19.991
The hem estimator is loud to sit for 1
to 2 minutes

00:03:19.991 --> 00:03:22.994
to allow the cells to settle for counting.

00:03:23.286 --> 00:03:26.289
The chamber is carefully placed
on a microscope for counting.
